20 ElektroPhysik
3.1.8 Stabilisation of calibration values.
No recalibration is necessary in changeable
external conditions, e.g. variations in ambient
temperature, as the gauge automatically takes these
into account.
3.2
Points to remember when calibrating
When calibrating according to sections 3.2.2 to 3.2.3,
the basic procedure is always as follows:
1.
Start calibration by pressing the appropriate
calibration key (ZERO or CAL).
2.
Apply the probe to calibration foil several
times.
3.
If necessary, adjust the value shown on the
LC display to the calibration foil value by
means of the arrow keys.
4.
End calibration by pressing the calibration
key again (ZERO or CAL).
Points to remeber when calibrating