22 ElektroPhysik
3.2.2 One-point calibration without foil
(zero only)
1.
Press ZERO to initialise ZERO calibration.
The display will show ZERO (flashing) and
MEAN (steady). MEAN indicates that the
mean value of the readings will be shown on
the display.
2.
Place the probe on the uncoated sample
(zero coating thickness) and raise it after the
bleep.
Place the probe on the uncoated sample
several times. The display always shows the
mean value of the previous readings.
To discontinue ZERO calibration, press
CLEAR.
3.
Press ZERO to end zero calibration. The word
ZERO will appear on the display (steady).
One-point calibratoin without foil