26 ElektroPhysik
Note:
If, for instance, an incorrect calibration value is
measured by mistake, calibration can be
discontinued by pressing CLEAR. The gauge
returns to the measuring mode using the previous
calibration.
6.
Adjust to the thickness of the foil with the
arrow keys.
7.
Press CAL to end CAL calibration. CAL will
appear on the display (steady).
8.
Now take readings by placing the probe on
the unknown coating and by raising it after
the bleep.
9.
It may be necessary to delete the CAL
calibration if, for example, an incorrect
calibration is entered. In this case:
a)
Press CAL then CLEAR to delete the CAL
calibration and any existing ZERO
calibration.
Two-point calibration