110
FOR SCIENTIFIC RESEARCH PURPOSES
| 8100550-03
•
2018-10-26 | GeoScan User Manual
Electromagnetic Immunity for Non-life-
supporting Equipment
Table A-3.
Electromagnetic compatibility (EMC) immunity guidelines and declarations for non-life-
supporting equipment (such as GeoScan components)
GeoScan systems are intended for use in the electromagnetic environment specified
below. The customer or user of a GeoScan system should ensure that it is used in such an
environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V rms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the GeoScan system,
including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation:
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800 MHz
d = 2.3
√
P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer, and d is the
recommended separation distance in
meters (m). Field strengths from fixed RF
transmitters, as determined by the
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may
occur in the vicinity of equipment
marked with the following symbol:
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is
Summary of Contents for Geoscan
Page 8: ...8 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 16: ...16 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 42: ...42 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 48: ...48 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 66: ...66 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 100: ...100 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 106: ...106 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 114: ...114 FOR SCIENTIFIC RESEARCH PURPOSES 8100550 03 2018 10 26 GeoScan User Manual ...
Page 115: ......