SE-1201 Electrocardiograph User Manual System Setup
- 90 -
Item
Description
Record Mode
Choose from:
Save Paper
or
Quickly
Select
Save Paper
,
10s after
pressing the
PRINT/STOP
key
on the
main
screen,
an ECG report is printed. The patient information, measure
information, interpretation and ECG waves are printed at the same time.
Select
Quickly
,
pressing the
PRINT/STOP
key
on the main screen
to
begin printing an ECG report immediately, the patient information,
measure information, interpretation and ECG waves are printed on the
different pieces of paper.
NOTE:
1. In the auto mode, only when
Sampling Mode
is set to
Real-time Sample
,
Quickly
is available.
2.
When
Record Style
is set to
3×4
,
3×4+1R
or
3×4+3R
, only
Save Paper
is available. When
Record Style
is set to
12×1,
only
Quickly
is available
Record Sequence
Choose from:
Sequential
or
Synchronous
Select
Sequential
, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the
previous lead group.
Select
Synchronous
, the lead group is printed one by one in a certain
sequence. All leads are printed with the same start time.
Gain
You can set the indicated height of 1mV ECG on the paper.
Choose from:
10mm/mV
,
20mm/mV
,
10/5mm/mV
,
AGC
,
2.5mm/mV
or
5mm/mV
.
AGC
means auto gain control. When ECG signals vary greatly,
AGC
can
be selected to adjust the gain automatically according to actual signals.
10/5mm/mV
means that the gain of limb leads is set to
10mm/mV
, while
the gain of chest leads is set to
5mm/mV
.
Record Device
Choose from:
Thermal
,
HP 2010/1050/2000
or
HP 2015/2035
You should connect the corresponding USB printer
HP 2010/1050/2000
and
HP 2015/2035
to the electrocardiograph.
WARNING
If the printer used is not the type listed above, additional safety measures (such as
applying an isolation transformer to supply the medical system) should be taken when the
safety of the medical system has not been evaluated. If in doubt, consult our technical
service department or your local distributor.