SE-1201 Electrocardiograph User Manual System Setup
- 97 -
10.6 Transmission Setup
NOTE:
1.
To transmit ECG data in DAT format to the PC, the Smart ECG Viewer software of the
manufacturer must be installed in the PC. You should log into the Smart ECG Viewer
software before transmission.
2.
To transmit ECG data in SCP/FDA-XML/PDF format to the PC, the FTP receiving
software must be installed in the PC. You should log into the FTP receiving software
before transmission.
Select
Transmission
on the
System Setup
screen, and press
Enter
to open the
Transmission
Setup
window.
Figure 10-7 Transmission Setup Window
Item
Description
Device No.
If you input
0
in the
Device No.
textbox, after you save data to the U disk
or SD card, the data will be in the directory of
ECGDATA\ECG-0\Store\Examination Date
of the U disk or SD card;
after you export files from the electrocardiograph to the U disk or SD
card, the files will be in the directory of
ECGDATA\ECG-0\Export\Export Date and Time
of the U disk or SD
card.
Auto Transmission
Choose from:
On
or
Off
Select
On
, ECG data will be transmitted automatically after an ECG
report is printed out in the auto or rhythm mode; in the off mode, 10s
ECG data sampled before pressing the
PRINT/STOP
key can be saved
and transmitted, but can not be printed.