SE-1201 Electrocardiograph User Manual Entering Patient Information
- 45 -
3.
When the main screen is displayed, scan the patient’s bar code with the bar code reader, and
then the patient information will appear in the corresponding box.
NOTE:
Only bar code readers, complying with the standards Bar Code 128
and
Bar Code
93 and recommended by the manufacturer, can be used.
5.3 Entering Patient Information by Acquiring Orders
NOTE:
To use the order function, the Smart ECG Viewer software of the manufacturer
must be installed in the PC.
Operation procedures are as follows:
1.
Connect the electrocardiograph to the PC with Ethernet cable recommended by the
manufacturer.
2.
Log into the Smart ECG Viewer software.
3.
Set
Remote IP
,
Local IP
,
Gateway
and
Subnet Mask
in the
Transmission Setup
window.
For details, please refer to Section 7.1.1 “Transmitting ECG Data in DAT Format through
Ethernet Cable”.
4.
Select
Order Acquired
function.
1)
Select
Setup> Patient Info
to open the
Patient Information Setup
window.
2)
Select
On
from the
Order Acquired
list box, and then press
Enter
to confirm.
5.
Select
Patient
on the main screen1 to open the
Patient Information
window.