SE-1201 Electrocardiograph User Manual Technical Specifications
- 121 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC/EN 60601-1+A1+A2, IEC/EN60601-1-2+A1,
IEC/EN60601-2-25, ANSI/AAMI EC11,
IEC/EN 60601-2-51
Anti-electric-shock type:
Class
І
with internal power supply
Anti-electric-shock degree:
CF type with defibrillation-proof
Degree of protection against
harmful ingress of water:
Ordinary equipment (Sealed equipment without liquid
proof)
Disinfection/sterilization
method:
Refer to the user manual for details
Degree of safety of
application in the presence of
flammable gas:
Equipment not suitable for use in the presence of
flammable gas
Working mode:
Continuous operation
EMC: Group
І
, Class A
NC <10
μ
A (AC) / <10
μ
A (DC)
Patient
Leakage
Current:
SFC <50
μ
A (AC) / <50
μ
A (DC)
NC <10
μ
A (AC) / <10
μ
A (DC)
Patient
Auxiliary
Current:
SFC <50
μ
A (AC) / <50
μ
A (DC)
A1.2 Environment Specifications
Transport & Storage
Working
Temperature: -20
℃
(-4ºF) ~ +55
℃
(+131ºF)
+5ºC (+41ºF) ~ +40ºC
(+104ºF)
Relative Humidity:
25%~93%
Non-Condensing
25%~80%
Non-Condensing
Atmospheric Pressure:
700hPa ~1060hPa
860hPa ~1060hPa