SE-1201 Electrocardiograph User Manual System Setup
- 106 -
Item
Description
External Output
The external output socket is equipped in the electrocardiograph, through
which the electrocardiograph can send signals to the external equipment.
Choose from:
Off,
Standard
or
Triggered
If
External Input
is set to
On
, and
External Output
is set to
Standard
or
Triggered
, the electrocardiograph sends the signals which it receives.
If
External Input
is set to
Off
, and
External Output
is set to
Standard
,
the electrocardiograph sends ECG signals of rhythm lead 1.
If
External Input
is set to
Off
, and
External Output
is set to
Triggered
,
the electrocardiograph sends pulses with the height of 5V and the width of
40ms, based on the data of rhythm lead 1.