SE-1201 Electrocardiograph User Manual Introduction
- 17 -
1)
Battery Compartment
Rated Voltage: 14.8V
Rated Capacity: 2200mAh
WARNING
1. Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the decrease of the battery capacity. Therefore, it is necessary to read the
user manual carefully and pay more attention to warning messages.
2. When leakage or foul smell is found, stop using the battery immediately. If your skin
or cloth comes into contact with the leakage liquid, cleanse it with clean water at once.
If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with
clean water first and go to see a doctor immediately.
3. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification must be used.
4. Only when the device is off can the battery be installed or removed.
NOTE:
If the battery has not been used for two months or more, you should recharge it
before using it again.
2)
Fuse
There are two fuses of the same specification installed on the bottom of the main unit. The
specification is: T3.15AH 250V Ø5×20.
WARNING
Ruptured fuses must only be replaced with those of the same type and rating as the
original.
2.6 Function Features
♦
Supporting AC and DC power supply modes, internal rechargeable li-ion battery with
professional battery powered circuit, battery management and protection systems
♦
Supporting multi-language
♦
Full alphanumeric keyboard (touch screen is optional)
♦
ECG signals of 12 leads are gathered and amplified simultaneously, 12-channel waves
are displayed and recorded simultaneously
♦
Correct detection for failure electrodes
♦
Convenient operation of recording by pressing the
PRINT/STOP
key with high
efficiency