SE-1201 Electrocardiograph User Manual Printing ECG Reports
- 51 -
2.
When the main screen is displayed, press the
MODE
key to select the R-R analysis
mode.
Press
F4
to switch the gain. Press
F5
to set the EMG filter or the Lowpass filter.
3.
Press the
PRINT/STOP
key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 180s, it begins to analyze and print an ECG
report.
4.
It will stop automatically after a complete R-R analysis report is printed, or press the
PRINT/STOP
key to stop printing the ECG report.
NOTE
: In the R-R analysis mode, you can not set the speed. The constant speed is
25mm/s and the printing speed is 5mm/s, because in the R-R analysis mode, the
ECG wave length is compressed to one fifth of the original wave length.
6.1.5 Review Printing
In the auto or off mode, after you press the
Review
key
,
10s ECG data sampled before you press
the key will be printed out.
NOTE:
The system will not respond to the
REVIEW
key unless 10s data has been
sampled on the main screen.
6.2 Copy Printing
In the auto and rhythm modes, pressing the
1mV/COPY
key can print the ECG report which was
printed out last time. Pressing the
PRINT/STOP
key can stop printing the ECG report.