Ser in u s ® 30 U ser M an u al 3 .3
Page 66
Time
[Backgrounds]
Edit and display the time that the next background will run. The time is
set using a 24-hour clock.
Repeat
[Backgrounds]
Defines an interval value for the repeat of the background based on the
Units
selected. This field indicates the delay period; once the specified
amount of time has lapsed the background will automatically run again.
The user can edit this field but some restrictions apply depending on the
Units selected. Default is “1”.
Units
[Backgrounds]
This is where the user can define the type of units for the
Repeat
delay
period. For example: A
Repeat
of
“
1
” and
Units
of “
Days
”
means that a
calibration will automatically run every day at the defined time. Default
is “Days”
.
3.4.10
Calibration Menu
Main Menu
→
Calibration Menu
Calibrating the instrument should be done with care (refer to Section 5 before using these menus).
Figure 52
–
Calibration Menu Screen
Cal. Type
Depending on the selection in this field, a number of extra menu items
will be displayed. These are separately documented in
Manual Mode
(refer to Section 3.4.10.1) and
Timed Mode
(refer to Section 3.4.10.2).
Select the
Cal. Type
field and select either
Timed
or
Manual
.
Timed
mode is an automatic calibration controlled by the:
▪
Interval between cycles
▪
Length of each calibration cycle
▪
Time when the calibration will begin
▪
Check only or automatic compensation
Timed calibration with span compensation enabled does not fulfil U.S.
EPA approval.
Manual
mode allows the user to choose the type of calibration they
wish to perform and will open the appropriate valves in preparation
for the user to perform a manual calibration. The setup used will
depend on the
Cal. Mode
selected. Manual mode is set as default.
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...