Communications
Page 107
Property
Description
Database Name
This is the name to be used for the table in the SQL database
containing this station’s data. It must be unique for each
station.
Device ID
Enter the Serial ID of the instrument. If the user is not using
multidrop;
this can be set to “0” or left blank.
Link Type
Select the type of connection used to connect to the
instrument. Different properties will appear depending on the
link type selected. Align these settings with those of the
instrument.
Log Interval
This needs to be the same as the
Data Log Interval
setting on
the instrument.
Note:
The available fields for communication parameters will change when the user changes the
link type. The user will need to set the communication parameters that have been defined on the
instrument.
5.
Once the station has been created, save the station by clicking the Save
shortcut icon or
File
→
Save.
6.
Click Acquire Configuration
.
This will probe the instrument for a channel list. After a few seconds,
the channel list should be visible in the
Channels
tab.
Figure 95
–
Station Configuration (Channel List)
Note:
If there was an error connecting to the instrument, a red dot will appear next to the station
name in the station list (on the far left hand side). Hovering over the red dot will present the user
with an error message (refer to Figure 96).
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...