Ser in u s ® 30 U ser M an u al 3 .3
Page 60
3.4.4
Status Menu
Main Menu
→
Analyser State Menu
→
Status Menu
The
Status Menu
presents a list of the current
Pass/Fail
status of the main components. During
warm-up, the status of some parameters will be a dashed line.
Figure 46
–
Status Menu Screen
Event Log
This field enters a screen with a log of all the events that the instrument
has performed. These events include errors and warnings. This log is
stored on the USB memory stick. The log is organised by month. When the
user enters this screen they will be prompted to enter the month for
which they wish to view events.
Show Error List
This field allows the user to display the list of current errors and warnings
on the screen.
Next Service Due
This field is visible with the next service due date if the service is due
within the next two weeks.
Maintenance Mode
Error if the system is “
In Maintenance
” (
refer to Section 3.4.13).
+5V Supply
Pass if the +5 V power supply is within the acceptable range.
+12V Supply
Pass if the +12 V power supply is within the acceptable range.
+Analog Supply
Pass if the analog power supply is within the acceptable range (+12 V).
-Analog Supply
Pass if the analog power supply is within the acceptable range (-12 V).
A2D
Fail only if a problem is detected with the analog to digital conversion.
Cell Temp.
Pass if the cell heater temperature is within ± 10% of the heater set point.
CO Scrubber Temp.
Pass if the CO scrubber heater temperature is within ± 10% of 90 °C.
Mirror Temp.
Pass if the mirror heater temperature is within ± 10% of set point.
Lamp/Source
Fail if the reference voltage is less than 0.5 V and the input pot is over 250,
which can be caused by the IR source failing.
Correlation Wheel
Fail if the correlation wheel is not functioning.
Ref Voltage
Checks that the reference voltage is within acceptable limits 3.6 - 4.4 V.
IR Source
Checks that the IR source voltage is 5 ± 0.5 V.
System Power
Pass if the system has an adequate electrical supply.
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...