Ser in u s ® 30 U ser M an u al 3 .3
Page 28
1.5.2
Sample Filter Holder
Refer to Figure 3 for the location of sample filter holder. Within the sample filter holder is a particulate
filter. The particulate filter is a Teflon 5-micron (µm) filter with a diameter of 47 mm. This filter prevents
all particles larger than 5 µm from entering the measurement system that could interfere with sample
measurement.
Note:
The 5-micron filter element should be replaced on a regular basis.
Figure 5
–
Sample Filter Holder
1.5.3
CO-CO
2
Converter
Refer to Figure 3 for the location of CO-CO
2
converter. The CO-CO
2
catalytic converter uses platinum
impregnated alumina beads heated to 90 °C as a conversion agent. The catalyst achieves conversion
of 0 - 200 ppm CO to less than 0.1 ppm CO, even in the presence of up to 2% water.
Within the body of the CO-CO
2
converter assembly, there is a heat exchanger used to precondition the
sample gas up to temperature. The sample gas is then drawn into the measurement cell.
Figure 6
–
CO-CO
2
Converter
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...