Calibration
Page 125
5.3
Manual Background
A background is a measurement that is made when CO free air is drawn through the reaction cell. The
resulting measurement signal (background) will include signals unrelated to CO. The background is an
electronic baseline, it is normally run every 24 hours automatically. There is sometimes a need for the
user to run a manual background.
Note:
A manual background is recommended before commencing a multipoint precision check.
Equipment Required
•
N/A
Procedure
1.
Open -
Main Menu
→
Calibration Menu
.
2.
Start -
Manual Background
→
Running
.
3.
Manual background will now display running and will proceed to draw air from the CO-CO
2
converter for 4 minutes (3 minutes of fill time and 1 minute of measure). The user can press stop
at any time to cancel the background. At the end of the 4 minute period a new background voltage
will be recorded in the event log as well as in the
Calculation Factors Menu
.
5.4
Zero Calibration
Zero calibrations are used to determine the zero response of the instrument and apply an offset to the
reading.
Performing a zero calibration will adjust the
Zero Offset CO
. This offset can be checked in the
Main
Menu
→
Service Menu
→
Calculation Factors Menu
and should be very close to zero. A large offset
may indicate a problem with the instrument (refer to Section 7).
Note:
Ecotech encourages regular zero precision checks; however Ecotech recommends that the
zero calibration only be performed when specifically required as it may mask issues that should be
addressed during maintenance/service.
A zero calibration can be performed either through the
Calibration Port, Background Air Port
or the
Sample Port
. Refer to the instructions outlined in the next three sections:
Note:
Ensure the instrument has been running for at least one hour before any calibration is
performed to ensure sufficient stability.
5.4.1
Calibration Port
Equipment Required
•
Zero Air Source (refer to Section for 1.3 definition)
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...