Communications
Page 101
Figure 86
–
Analog & Digital I/O Individual Pin Descriptions
CAUTION
The analog and digital inputs and outputs are rated to CAT I.
Exceeding 12 VDC or drawing greater than 400 mA on a single output or a total
greater than 2 A across the eight outputs can permanently damage the
instrument and void the warranty.
4.5
Logging Data
When the user receives the instrument from the factory it will have a default set of parameters already
setup in the internal data logger. These select few parameters have been chosen for their relevance in
assisting in troubleshooting the instrument.
4.5.1
Configure Instrument Internal Logging
In order to log data the user must first specify a data logging interval. This is how often data will be
logged to the USB memory stick. The user can log a maximum of 12 parameters. These parameters can
be selected by name or by parameter number using the parameter list as a reference (refer to
Table 33).
4.5.1.1
Data Log Setup
–
Numeric
Procedure
1.
Open -
Main Menu
→
Communications Menu
→
Data Logging Menu
(refer to Section 3.4.23).
2.
Select -
Data Log Interval -
(adjust to the desired value) - Accept.
3.
Open -
Data Log Setup -Numeric -
(select the parameter numbers user wishes to log).
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...