Service
Page 155
Figure 141
–
Test Point Location
6.5
Bootloader
The Serinus Bootloader is the initial set of operations that the instruments’ microprocessor performs
when first powered up (similar to the BIOS found in a personal computer). This occurs every time the
instrument is powered up or during instrument resets. Once the instrument boots up, it will
automatically load the instruments’ firmware. A service technician may need t
o enter the Bootloader
to perform advanced microprocessor functions as described below.
To enter the Bootloader turn OFF the power to the instrument. Press and hold the plus key while
turning the power ON. Hold the Plus key until the following screen appears.
Ecotech Analyser / Calibrator
V3.0 Bootloader
Press ‘1’ to enter Bootloader
If the instrument displays the normal start up screen, the power will need to be toggled and another
attempt will need to be made to enter the Bootloader. Once successful, press 1 to enter the
Bootloader Menu
.
6.5.1
Display Help Screen
Once in the Bootloader screen it is possible to redisplay the help screen by pressing 1 on the keypad.
6.5.2
Communications Port Test
This test is very useful for fault finding communication issues. It allows a communication test to be
carried out independent to any user settings or firmware revisions.
This command forces the following communication ports to output a string of characters: Serial Port
RS232 #1, USB rear and Ethernet Port. The default baud rate is 38400 for the RS232 Serial Port. To
initiate the test press the number 2 key from the Bootloader screen.
6.5.3
Updating Firmware
It is important for optimal performance of the instrument that the latest firmware is loaded. The latest
firmware can be obtained by visiting Ecotech’s website:
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...