Ser in u s ® 30 U ser M an u al 3 .3
Page 36
Analog Inputs
The instrument is also equipped with three analog voltage inputs (0 - 5 VDC CAT 1) with resolution of
15 bits plus polarity.
CAUTION
Exceeding these voltages can permanently damage the instrument and void the
warranty.
Digital Status Inputs
The instrument is equipped with eight logic level inputs (0 - 5 VDC CAT 1) for the external control of
zero/span calibration sequences.
CAUTION
Exceeding these voltages can permanently damage the instrument and void the
warranty.
Digital Status Outputs
The instrument is equipped with eight open collector outputs, which will convey instrument status
conditions and warning alarms such as no flow, sample mode, etc.
CAUTION
Exceeding 12 VDC or drawing greater than 400 mA on a single output or a total
greater than 2 A across the eight outputs can permanently damage the
instrument and void the warranty.
Bluetooth
This allows for remote access of the instrument to any Android device with the
Serinus Remote
Application
installed on it. It uses Bluetooth to control the instrument, view parameters, download
data and construct real-time graphs.
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...