Calibration
Page 127
2.
Open -
Main Menu
→
Calibration Menu
.
3.
Select -
Cal. Type
→
Manual
- Accept.
4.
Select -
Cal. Mode
→
Measure
- Accept.
5.
Select -
Zero Source
→
External
- Accept.
6.
Allow the instrument time to achieve a stable response.
7.
Enter -
Zero Calibrate CO
- (read note) - OK.
8.
Disconnect the zero air source and reconnect the sample line to the
Sample Port
.
5.4.3
Background Air Port
Equipment Required
•
N/A
Procedure
1.
Open -
Main Menu
→
Calibration Menu
.
2.
Select -
Cal. Type
→
Manual
- Accept.
3.
Select -
Cal. Mode
→
Zero
- Accept.
4.
Select -
Zero Source
→
Internal
- Accept.
5.
Allow the instrument time to achieve a stable response.
6.
Enter -
Zero Calibrate CO
- (read note) - OK.
7.
Select -
Cal. Mode
→
Measure
- Accept (To return to sample measure).
5.5
Span Calibration
A span calibration is a calibration performed at the upper end of the instrument
’s measurement range.
Ecotech recommends calibration at 80% of the full scale measurement or operating range of the
instrument.
While the instrument range is commonly set as a default 0 - 50 ppm, this is widely recognised as no
longer being valid with modern digital communication and most regulators will now recommend a
range more suited to local conditions.
A span gas can be supplied through either the
Calibration Port
or
Sample Port
. Refer to the instructions
outlined in the next two sections.
Note:
Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure sufficient stability.
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...