Appendix
Page 205
Example
A request with a command of 4 and a message field of 85,64,64,0,0 would place the instrument into
Span mode.
Set Calibration
Enters a new calibration value: the same as entering Span Calibrate or Zero Calibrate on the calibration
menu.
Where:
Command byte 18
Message byte 1 0, 1, 2, or 3 where
0 = Span
1 = Zero (first zero gas)
2 = Zero (second zero gas)
3 = Zero (Third zero gas)
Message byte 2 - 5 The IEEE representation of the calibration value.
A.3
List of Parameters
Note:
Parameters are for all Serinus series analysers and may not be applicable to an
individual instrument.
Table 33
–
Advanced Protocol Parameter List
#
Description
Notes
1
Cal/Zero Valve
0 = Zero, 1 = Cal
2
Internal Span Valve
0 = OFF, 1 = ON
3
+Analog Supply
Positive analog supply voltage
4
Gas 5 Avg.
Average of the readings (for Gas5) of the last n minutes where
n is the averaging period E.g. Nx
5
Pregain
S30H linearization coefficient gain
6
Sample/Cal Valve
0 = Sample, 1 = Cal/Zero
7
NOx Measure Valve
0 = NO, 1 = NOx
8
NOx Bypass Valve
0 = NO, 1 = NOx
9
NOx Backgnd Valve
0 = OFF, 1 = ON
10
Valve Sequencing
0 = OFF, 1 = ON
11
LCD Contrast Pot
0 = Lightest, 255 = Darkest
12
SO2 Ref Zero Pot
S50 Reference zero pot
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...