Operation
Page 67
Zero Source
Select whether the instrument will sample from the external
calibration port or from the internal zero source when zero gas is
requested.
Cycle Time
The duration of each
Cal. Mode
(span and zero) when performing
Cycle Mode
(refer to Section 3.4.10.1) or
Cal. Type
is set to
Timed
.
(refer to Section 3.4.10.2)
Span Calibrate CO
This field is used to perform a span calibration and should be only used
when a known concentration of span gas is running through the
measurement cell and the reading is stable.
Activating the span calibrate field for a named gas will open a dialog
box. Enter the concentration of the span gas that the instrument is
sampling and press
Accept
.
Zero Calibrate CO
This command is used to correct the zero calibration setting. This
option should be used only when zero gas is running through the
measurement cell (refer to Section 5.4 before using this command).
Manual Background
Immediately perform a background.
Pressure Calibration Menu
Refer to Section 3.4.11.
Flow Calibration Menu
[Internal Pump Option]
Refer to Section 3.4.12.
Pressure CO
This field displays the cell pressure measured during the last
calibration.
Temperature
Cell temperature when the last span calibration was performed.
3.4.10.1
Manual Mode
These items appear in the
Calibration Menu
when
Cal. Type
is set to
Manual
.
Cal. Mode
When calibration type is set to
Manual
the instruments operational
mode can be chosen from the following:
Measure
: Is the normal measurement through the sample port.
Zero
: This mode will take air through the calibration port so that a zero
calibration can be performed. Data is flagged as zero data.
Span
: This mode will take air through the calibration port so that a span
calibration can be performed. Data is flagged as span data.
Cycle
: Performs a zero and a span calibration mode and returns to
measure mode. The length of time spent measuring each calibration
mode is set in
Cycle Time
(refer to Section 3.4.10).
While the instrument is still in the warm-up period (refer to Section 3.1)
the
Cal. Mode
cannot be changed from
Measure
mode.
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...