Ser in u s ® 30 U ser M an u al 3 .3
Page 68
3.4.10.2
Timed Mode
These items appear in the
Calibration Menu
when
Cal. Type
is set to
Timed
.
Date
Enter the date for the next calibration to start.
Time
Enter the time that calibration will be performed. The time is set using a
24-hour clock.
Repeat
The calibration will automatically run again after the specified amount of
time. This field specifies the delay period (from 1 - 20,000 units, as
specified below).
Units
This is where the user can define the type of units for the
Repeat
delay
period. For example: A
Repeat
of
“
3
” and units of “
Days
”
means that a
calibration will automatically be performed every three days.
Span Compensation
When
Enabled
the instrument will automatically perform a
Span
Calibrate CO
at the end of the
Cycle Time
and adjust the gain based on
the
Span Level
.
When
Disabled
it will do a precision check only, no adjustment is made.
Timed calibration with span compensation enabled does not fulfil U.S.
EPA approval or EN certification.
Span Level
Enter the concentration of span gas expected. Used when the
Span
Compensation
is
Enabled.
3.4.11
Pressure Calibration Menu
Main Menu
→
Calibration Menu
→
Pressure Calibration Menu
Entering this menu will set the valves to the pressure calibration configuration; leaving the menu will
restore the valves to normal operation (refer to Section 5.2).
Figure 53
–
Pressure Calibration Menu Screen
Vacuum Set Pt.
The zero point for the calibration. Activating this item will open a
dialog box of instructions.
Ambient Set Pt.
The high point for the calibration. Activating this item will open a
dialog box of instructions.
Pressure Units
Select the units that the pressure will be displayed in (
torr
,
PSI
,
mBar
,
ATM
or
kPa
).
Summary of Contents for Acoem Serinus 30
Page 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Page 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Page 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Page 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Page 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Page 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Page 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Page 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Page 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Page 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Page 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Page 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Page 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Page 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Page 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Page 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Page 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Page 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...