417101637_Lub-O-Mat
- 45 -
Rev. 03-07.12
9.
Troubleshooting
CAUTION
Repair work on lines that are live or under pressure should be carried out only by qualified
personnel.
Symptom
Cause/Fault
Response
Metering pump not
operating; no green
LED display
Incorrect voltage
Confirm that supply voltage is being supplied;
exchange any faulty fuses in the terminal box
or pump. (See EMP II Metering Pump
Operating Guide)
Pump not metering
Pump faulty
Refer Elados EMP II Metering Pump
Operating Guide.
Pump incorrectly set
Set correct pulse divisor or multiplier.
No pulse transfer from water flow
sensor to EMP pump
Check plug on pulse transfer cable.
Replace water flow sensor if necessary.
Fault warning on
display of the pump
Metering valve clogged
Clean valve
Metering back pressure to high
Check metering line
Red LED lights up on
“empty” signal
indicator AND Pump
not metering
Supply container empty
Replace supply container
Suction pipe faulty
Replace faulty suction pipe and if necessary
check whether the float switch is blocked.
No water flow
Fresh water armature damaged
Check fresh water armature, in case remove
Summary of Contents for 183042
Page 14: ...417101637_Lub O Mat 14 Rev 03 07 12 5 8 Stromlaufplan Abb 5 8...
Page 24: ...417101637_Lub O Mat 24 Rev 03 07 12 11 Konformit tserkl rung...
Page 37: ...417101637_Lub O Mat 37 Rev 03 07 12 5 8 Circuit diagram Fig 5 8...
Page 47: ...417101637_Lub O Mat 47 Rev 03 07 12 11 Declaration of conformity...
Page 60: ...417101637_Lub O Mat 60 Rev 03 07 12 5 8 Sch ma de circuit fIG 5 8...
Page 70: ...417101637_Lub O Mat 70 Rev 03 07 12 11 D claration de conformit...
Page 71: ......