3.
Connecting to a Target System and Connector
3.1. Signal List
The following table shows the signal list of the target end of the target probe connector when
you use FR830.
Microcomputer
Signal
NET IMPRESS Standard Signal
Microcomputer
Signal
EVSS
GND
15
1
GND
EVSS
TVpp1
16
2
TVccd
EVCC
TVpp2
17
3
Vcc
Watchdog pulse
WDC
(18)
(4)
TRES
Positive/TRES
TAUX3
TVpp1c
19
5
/TRES
/RESET
FLMD0
TAUX4
TVpp2c
20
(6)
TCK
FPCK
(JP0_2)
reserved
21
7
reserved
reserved
22
8
reserved
TAUX
23
9
TAUX2
(TRW)
TBUSY
24
(10)
/TICS
Multiplexer
TI0
25
11
TAUX5
(/TOE)
TVccs
26
(12)
TMODE
FLMD1
FPDT
(JP0_1)
TRXD
(27)
13
TTXD
FPDR
(JP0_0)
EVSS
GND
28
14
GND
EVSS
Target Probe Signal List (FR830)
1. Be sure to connect all the signals marked with a circle “○ “ to the target system.
2. For the signals in parentheses, connect them only when necessary.
3. For the signals not marked with “
〇
“ or not in parentheses, be sure not to connect them to a
circuit of the target system.
For inquiry about the detail information, please contact our Support Center.