Table 10-6 millivolts (source) error limits
Source
(mV)
Calibrator
uncertainty
(mV)
Permitted
DPI620G-IS
error (mV)
0 0.0001
±0.0060
100 0.00046 ±0.0090
195 0.0009 ±0.0119
1000 0.003 ±0.0750
1995 0.006 ±0.0899
Table 10-7 Voltage (source) error limits
Source
(V)
Calibrator
uncertainty
(V)
Permitted
DPI620G-IS
error (V)
0 0.000004
±0.00048
3 0.000019
±0.00060
6 0.000034
±0.00072
9 0.000049
±0.00084
12 0.000064
±0.00096