ERROR
TYPE
– The deviation from which the
transfer characteristic is to be
calculated. This can be one of the
following:
%
Span
a percentage of the Input signal span.
%
Rdg
a percentage of the Input signal reading.
•
LINEARITY
- The transfer characteristic from
Reference to Input signal. This can be
one of the following:
Linear
a proportional response.
Square
Root
often found in flow sensors.
•
TOLERANCE
– The test limits for the deviation from the
transfer characteristic.