User Manual
Druck DPI620G-IS
22 Feb 2017
116M5464 Revision A
Page 56 of 248
2.7
Example Procedure: Measure or Source Current with
Internal 15V Loop Power, CH2
Note:
CH2 can be configured to measure or source current
using internal or external loop power. For an example of
external loop power, refer to
Section 2.6.2
.
Safety Note:
Internal 15V loop power is activated for CH2
functions measure and source current with internal 15V
loop power. With these set-ups an unregulated 15V supply
is presented across terminals ‘15Vo’ and ‘mA-/COM’ as
shown in Figure 2-15
.
Figure 2-15 Connection Details for 15Vo Unregulated Current
Loop Output