Table 10-4
or Table 10-5
)
.
Table 10-4
Millivolts (measure) error limits
Applied
(mV)
Calibrator
uncertainty
(mV)
Permitted
DPI620G-IS
error (mV)
±1995
0.051
±0.0899
±1000
0.040
±0.0750
±195
0.0051
±0.0109
±100
0.0040
±0.0095
0
(short circuit)
0.0036 ±0.0080
Table 10-5 Voltage (measure) error limits
Applied
(V)
Calibrator
uncertainty
(V)
Permitted
DPI620G-IS
error (V)
±30
0.00052 ±0.00195
±21
0.00040 ±0.00172
±20
0.00031 ±0.00090
±10
0.00016 ±0.00065
±5 0.00008
±0.00052
0
(short circuit)
0.000024 ±0.00040