5.3
Analysis Function
Set Input channel parameters (refer to section 5.2), and
return to
CHANNEL
SETUP
screen.
Select the Start button
The Analysis window displays the following:
•
The deviation of each Input channel from the ideal
transfer characteristic.
•
A tolerance limit test icon
Pass
(within tolerance test limit)
Fail
(outside tolerance test limit)
To check the full range of the device do the following:
1.
Step the Reference signal value through its range.
2.
Check the Analysis window at each step.
3.
If the Reference is sourced from the calibrator, move to
the channel window to change the Reference value.
4.
Return to the Analysis window.
5.
When the analysis is complete exit the window by
selecting exit .