In order to put the calibrator into
mode, the procedure on the previous page
should be followed.
CALIBRATION
The following is a more detailed explanaon of point 2 „2. Preparaon for
“
CALIBRATION
a) Number of the calibraon and therefore the test item (up to 16 calibraons, each with up to
32 test points can be predefined and then recorded).
b) Test item type and respecve test item measuring signal: [
/
/
/
/
0-20 mA 4-20 mA 0-1 V 0-2 V
0-5 V 0-10 V
mechan.
/
or.
for analogue dial pressure gauges].
c)
no.: Ident number of the test item
ID
d) Tag no.:
number of the test item
TAG
e) R.-Start/Range-End:
Start of measuring range and end of measuring range of the currently to calibrate test item
f) Unit and resoluon (submenu)
g) Measurement uncertainty of the test item in %
(i.e. of the span) or % rd (i.e. of the
FS
measured value)
h) Measurement type for the test item (relave/gauge or absolute)
i) Test medium (pneumac -> „gas“ or hydraulic -> „oil“)
j) Voltage supply for the test item (on/off).
If no external supply is required for the test item,
„
“ should be selected to conserve engergy.
OFF
k) Number of the test point x
l) Oponal delay me [sec] (see under chapter 7.3.4 „
mode (preparing the test
CALIBRATION
points of a calibraon)“)
m) Value of the test point x (input via numeric keypad) (test point x+1 or x-1 accessible via
cursor-down-arrow or cursor-up-arrow keys)
n) True value of the test item (will be recorded during the calibraon; if test item is a pointer
pressure gauge, the measured value has to be entered via numeric keypad)
bar
psi
MPa
mH2O
kg/cm²
User: 1.00000
Resolution: 10,000
mbar
inHg
kPa
cmH2O
inH2O
hPa
cmHg
Pa
mmH2O
mmHg
Unit of test item
...Unit/Resolution
Select and confirm
(standard units) via
Customer-specific unit; with
respect to bar (input via
numeric keypad, confirm
with
key
ENTER
Display resoluon in operang mode
(back with )
BACK