LPC 300 Druckkalibrator
300 Pressure Calibrator
LPC
94
In order to switch the calibrator into
mode, follow the instrucons on the previous
page.
MEASURING
The following is a more detailled explanaon of point „2. Preparaon for
“
MEASURING
a) Test item type and test item measurement signal: [----] for measurement without test item.
b) R.-Start / Range-End:
Not inputs or selecons here, as not necessary for measurement without test item.
c) Unit and resoluon (submenu): No inputs or selecons here, as not necessary for
measurement without test item.
d) Class:
Not inputs or selecons here, as not necessary for measurement without test item.
e) Pressuretype:
No inputs or selecons here, as not necessary for measurement without test item.
f) Medium:
Test medium (pneumac -> „gas“ or hydraulic -> „oil“)
g) PowerSupply:
Voltage supply for test item (on/off).
As no test item is connected, „
“ should be selected to conserve energy.
OFF
Inverse: current cursor posion, alter via , confirm with
xxx
vw tu
Parameter selecon from list or menu via , confirm with
0.000
Parameter input via numeric keypad, confirm with
CLEAR
Clear entry with