With this example, the definion of individual test points / pressure steps for a calibraon is
clarified. It is possible to prepare up to 16 calibraons, each with up to 32 test points.
• Accessing the menu points:
Via
key and selecon of the menu point:
SETUP
CALIBRATION
(see chapter 7.3.3 „
mode“)
CALIBRATION
Enter the desired test points in the way described on the previous page.
With the calibraon of pressure measuring instruments with electrical output signals (pressure
transmi ers) the reference is calibrated to the display (i.e. the pressure is always adjusted so that
the reference value matches the setpoint exactly). Since in all cases, an exact se ng of the
pressure is not always possible, along with the actual value (test item value) and the setpoint
(reference value), the true reference value is also recorded.
With the oponal
soware one of the two combinaons can be listed in the calibraon
cerficate.
LPC-
Cal
If the calibraon is to follow
-/
kkS guidelines, then the measured value for
DKD DA
the subsequent test point should not be recorded unl a defined me has passed
(for example 30 sec.), consisng of a load change me and se ng me (see below
figure: A Calibraon cycle to
/
kkS direcve 6-1 for measurement
DKD DA
uncertainty > 0.6% of measuring span.
oponal
[se ]
delay me
c
Input
c keypad
via numeri
a
confirmaon with
.
nd
ENTER