b) R.-Start/Range-End:
Start of measuring range and end of measuring range of the currently connected
LPC-S
reference pressure sensor.
c) Unit:
Basic pressure unit of the
reference pressure sensor.
LPC-S
d) Class:
Accuracy of the measuring chain of the
with connected
reference
LPC
LPC
300
-S
LR-
Cal
pressure sensor.
e) Pressuretype:
Pressure type of the currently connected
reference pressure sensor
LPC-S
(Gauge/relave pressure or absolute pressure)
f) Overload:
Length of me for which the
reference pressure sensor was unacceptably overloaded.
LPC-S
If the value here is not equal to zero, then it is highly probable that the calibrator
no longer meets its specified accuracy class. The only soluon for this is an
immediate recalibraon. (For absolute pressure sensors <1 bar this funcon
is deacvated, since for this measuring range atmospheric pressure already
represents an overload.)
g) Ref-Value:
Current measured value of the connected
reference pressure sensor.
LPC-S
h) Offset:
This menu opon only appears if the
reference pressure sensor is an absolute pressure
LPC-S
sensor. Through this menu opon the measured value of the reference pressure sensor can
be adjusted.
This should only be used, however, as close as possible to the absolute zero, and using a
reference that is at least 4 mes more accurate than ±0.025%
.
FS
i) Cal-Dat.:
Calibraon date of the
reference pressure sensor (
/
/
).
LPC-S
YYYY MM DD