LPC 300 Druckkalibrator
300 Pressure Calibrator
LPC
113
Faults
Causes
Measures
An intermi end audio alarm is
audible and the display shows
lines rather than the measured
value.
The current pressure value is
more than 10% outside the
allowable pressure range.
Adjust the pressure
immediately and
appropriately.
The
reference pressure
sensor is not properly
connected, or there is a
problem with the connecon.
LPC-S
Connect the
reference
pressure sensor properly.
Check connecon between
and
.
LPC-S
LPC
LPC
300
-S
LR-
Cal
The test item (read via mA or
V measuring input) is not
responding.
Test item is not properly
connected.
Check the cabling.
Test item is without its own
power supply.
Acvate the 24 V test item
power supply.
The test-item (read via mA or
V measuring input) displays a
false value.
Wrong measuring range
entered for the test item.
Enter correct measuring range
of the test item.
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the
and related
sensor(s) must be shut down immediately, and
LPC
LPC
300
-S
LR-
Cal
it must be ensured that pressure and/or signal are no longer present, and it must
be prevented from beeing inadvertently put back into service.
In this case, contact the manufacturer.
If a return is needed, please follow the instrucons given in chapter 10.2 „Return“.
10. Dismounng, return and disposal
WARNING!
Residual media in the dismounted
reference pressure sensor can result in
LPC-S
a risk to persons, the environment and the equipment. Take sufficient
precauonary measures.
10.1 Dismounng
Only disconnect the test and calibraon installaons once the system has been depressurised!