For gauge or relave pressure sensors, there is a pressure-equalising vent in the
top of the sensor under the black plasc cover. This vent (with integrated
membrane) must always remain clear!
CAUTION!
Only ever use the original connecon cable from
&
DRUCK
TEMPERATUR
Leitenberger. Only ever use those
reference pressure sensors, which are
LPC-S
cerfied together with your
calibrator. Otherwise, the calibraon
300
LPC
LR-
Cal
coefficients, necessary for the accuracy, are not stored in the
300
LPC
LR-
Cal
calibrator. The calibrator can handle up to 10 different
reference pressure
LPC-S
sensors.
CAUTION!
Pressure reference sensors
with pressure ranges up to 1000 bar can be
LPC-S
direct mounted at the
calibrator. This connecon can be extended
300
LPC
LR-
Cal
with the oponal extension cable order-code
.
LPC KABEL
-
Reference pressure sensors with pressure ranges > 1000 bar cannot be mounted
straight to the calibrator. Here you need in every case the extension cable with
order-code
.
LPC KABEL
-
4.4.2 Electrical connecon of the reference pressure sensor
-S to the
300
LPC
LPC
LR-
Cal
a) Standard electrical connecon (all pressure ranges up to 1,000 bar)
The
reference pressure sensor is connected electrically to the calibrator via a M12 x 1.5
circular connector with screw-fastening.
LPC-S