Dosper M/5
19
18
Remove the lock caps from the
coupling plugs (no. 19). Press the
quick-coupling/bottle adapter of the
silicone containers onto the coupling
plugs of the unit and let them snap
hearably.
Attention!
The components A and B
have to be connected according to the
printing on the unit and the applied
colour marking. Cut an opening of
approx. 0,5 cm
2
into the bottoms of
the cubitainers in order to let air come
in when producing. Remove protection
cap from mixing head (no. 4) and do
not reuse. Turn on the unit at the
operation button. Put a reception
bowl under the mixing head and
press push-button (no. 2). Keep the
push-button pressed until the test
silicone has been produced completely
and both components come out
homogeneously at the mixing head.
Fix mixing canula at the mixing head
(no. 4) by means of turning it to the
right by 90°. The turning of the
mixing canula can be eased by
means of the enclosed star-turn
handle (Item-no. 1565E).
The Dosper M/5 is now ready to ope-
rate. Put duplication flask under the
opening of the mixing canula and
press push-button (no. 2). As long as
the material is not cured in the mixing
canula (approx. 4 minutes) several
flasks can be filled in succession.
After having finished the filling
process do not remove the mixing
canula. Do only remove the mixing
canula before the next duplication
by means of a 90° turn to the right.
Attention:
Never produce against a
mixing canula with cured material!
Change of bottles/cubitainers
Remove the quick-coupling/bottle
adapter from the unit by means of
tearing upwards the lock-up. Screw
quick-coupling/bottle adapter onto
the new bottles/cubitainers and
connect as described above. Observe
the correct coordination of the com-
ponents according to the colour.
Observe the contents of the contai-
ners at any time! They must not be
emptied completely, as otherwise air
gets into the system and the unit has
to be bleeded under great effort.
A minimal filling of 2-3 cm has to
remain in the containers.
Care and maintenance
Disconnect the unit from mains
supply before care and maintenance!
Cleaning is effected with a mild
cleaning liquid.
Wipe away silicone remainders from
the unit. Keep coupling plug and
mixing head clean. The cleaning
liquid must not get into the unit.
General notes
Repairs on the Dosper M/5 may only
be executed by experts. Disconnect
the unit from mains supply at any rate.
Dreve company is only responsible for
the safety, reliability and performance
of the unit, if
- the unit is exclusively used for the
corresponding field of application.
- operation is only carried out by
accordingly trained personnel.
- the unit is operated in accordance
with these working instructions.
- extensions, new adjustments, changes
or repairs are executed by persons
authorised by the company.
As we are currently developing our
products, we reserve the right for
technical changes.