User Manual of CIC-D100 Ion Chromatograph
24
This setting defaults to 1000ms and does not need to be modified.
※
Collection interval
This parameter is related to spectrogram collection frequency. It defaults to 100ms and does
not need to be modified.
b Spectrogram software setting
Set the full installation path file name of the chromatography workstation software to
automatically start the spectrogram software when the chromatography workstation software is
not started.
Figure 3-4 Spectrogram software setting
※
Collection time
If the sampling collection time exceeds this set time, spectrogram collection will be
automatically stopped. After item [Injection Signal Triggers New Spectrum] is checked, the
sampling signal enables the spectrum triggering to start, otherwise the trigger signal will not be
responded.
※
Communication channel
This selection corresponds to the data receiving channel of chromatography workstation
software. The reverse control software will control and send data to the corresponding channel
window.