User Manual of CIC-D100 Ion Chromatograph
27
Figure 3-8 Equipment startup setting interface
This interface is to configure the startup sequence of each equipment. Select the items in the
left list, and click
button in the middle to move the selected items to the right.
In the startup sequence setting list, the delay command which is customizable can be added
in the middle. Click the
button on the left to pop up the window for adding new items as
shown below. Select the item "Delay [x]ms", and the input box will automatically add this item.
Change the X in brackets to the delay time to be set. The unit is ms.
For example: a delay of 1 second is set as delay [1000]ms
Figure 3-9 Delay setting interface
Pump flow rate item setting: pump flow rate setting can be inserted in the startup process
to increase the speed slowly.