User Manual of CIC-D100 Ion Chromatograph
61
B. Installation
B.1 Equipment requirements
1. Ambient temperature: 5
℃
to 30
℃
2. Relative humidity: relative humidity of 5% to 85%, non-condensing
3. The room should be clean and well ventilated. The instrument should be stably placed on
the workbench without strong mechanical vibration and electromagnetic interference
sources.
4. The instrument should be well grounded.
B.2 Unpacking
Unpack the instrument and check whether the equipment and accessories are damaged
during transportation. If damaged, the carrier company should be provided with a statement of
responsibility in time.
Check and count the articles against the packing list. If there are any omissions or errors,
please contact us in time.
:
since the instrument is heavy, please ask two or more people to carry the
instrument together and lift the bases on the left and right sides of the instrument. It is
forbidden to lift the front side and rear side, otherwise the soundness of the front door will
be damaged.
:
Do not contact sharp objects to avoid scratching the surface coating of the
instrument.
:
the power supply current used by this instrument is large, which may cause
personal injury such as electric shock. Therefore, please turn off the power supply and
unplug the power cord before handling. When handling the instrument, please fasten the
Caution
Caution
Electric
shock