User Manual of CIC-D100 Ion Chromatograph
18
Figure 2-7 Schematic diagram of exhaust port
High pressure limit
The pressure resistance of the eluent generator is ≤17MPa. To ensure the long-term stable
operation of the equipment, it is recommended that the daily working pressure is <15 MPa. If the
system pressure is too high, please carefully check the cause of overpressure in the flow path and
eliminate the problem in time to prevent damage to the device.
Low pressure limit
The device generates a large amount of gas during electrolysis. The gas needs to pass
through the degassing membrane under a certain osmotic pressure for elimination. The
recommended daily working pressure is >8MP.
Precautions
:
the liquid storage tank is filled with high concentration KOH or MSA
solution which is extremely corrosive. Please do not disassemble by yourself to prevent
injury accidents.
:
please confirm whether the system pressure is within the applicable range
(≤17MPa) before using this equipment to avoid damaging this equipment.
:
Before using the equipment, first start the infusion pump to flush the pipeline
with pure water for 5 minutes. Then, open the exhaust valve, set the concentration value
and power on; after using this equipment, close the exhaust valve, power off and flush the
pipeline with pure water for at least 10minutes.
Exhaust port
Warning
Warning