User Manual of CIC-D100 Ion Chromatograph
68
injection;
3. The suppressor waste liquid pipe discharges the waste liquid finally generated in the flow
path;
4. The waste liquid pipe of the eluent generator discharges the gas generated in the eluent
generator tank.
:
To prevent siphoning of waste liquid, please check whether the waste liquid
pipe is bent, squeezed or raised at any time.
B.8.1 Installation of waste liquid pipes
1. Connect all pipelines of the instrument;
2. Collect all waste liquid pipelines into a thicker pipeline;
3. Put the waste liquid collection pipe into the waste liquid bottle.
:
The waste liquid bottle should not be sealed because the continuous
self-regeneration suppressor uses electrolysis to suppress the background during the
operation of the instrument and a small amount of oxygen and hydrogen will be generated
during electrolysis. Gas should be prevented from staying in the waste liquid bottle so as
not to generate static electricity. In serious cases, there may be explosion.
B.9 Installation of eluent bottle
1. Rinse the eluent bottle with ultrapure water;
2. Load the eluent into a eluent bottle;
3. Place the eluent bottle in the eluent tray at the upper end of CIC-D100;
4. After installing the filter head, put the pipeline into the eluent bottle and tighten the eluent
bottle cap.
Caution
Caution