Intended Use
Dräger-Vapor
19.n – an anaesthetic vaporiser for
accurately-controlled enrichment of fresh gas with the
vapour of the relevant liquid anaesthetic agent: Isoflurane
or Halothane or Enflurane or Sevoflurane.
Only fill Vapor with the anaesthetic agent for which it
is designed.
If the wrong anaesthetic agent is used the patient may be
harmed.
The Vapor is inserted into the fresh gas line leading to the
breathing system of an inhalation anaesthetic machine
which has a continuous flow of fresh gas.
Do not connect Vapor downstream of the fresh gas
outlet on an anaesthetic machine.
Do not use Vapor in the breathing system.
Use only dry, pure medical gases.
Before connecting a Vapor
19.n to other manufacturers'
anaesthetic machines or operating it outside the range of
uses specified in the "Technical Data" it is essential to
consult Dräger Medizintechnik GmbH to check whether
the connection parameters required for it to function
properly (for instance, geometry, pressure and flow) will
be met.
Any deviations from the connection parameters
required may result in an incorrect concentration
being delivered.
Handle Vapor with care. Do not drop.
Do not carry by the handwheel or locking lever.
Vapor models:
The model numbers 19.1, 19.2 and 19.3 which apply to
otherwise identical machines refer to different types of
connections:
19.1 connection without Interlock
19.2 connection to Interlock 1
19.3 connection to Interlock 2