Vapor without Interlock
If two Vapors are connected in series without an Interlock
system or Vapor switch there is a risk that both Vapors
will be used at the same time. If this happens, gas
containing anaesthetic agent will flow from one Vapor
into the vaporising chamber of the other Vapor where the
two anaesthetic agents will mix – the mixture of
anaesthetic agents cannot be controlled.
Make sure that only one Vapor can be used at
any one time.
Dräger recommends that an Interlock system or a Vapor
switch is used without fail so that there is two-way
locking if two Vapors are connected.
When using two Vapors with ISO cones or hose
connections:
●
Never connect Vapors in series.
0
0
0
0
0
0
15
Preparation
Connecting Vapor