Filling volume for anaesthetic agent
about 200 mL with dry wick
about 140 mL with moist wick
about 135 mL between minimum and maximum mark
Consumption of anaesthetic agent [mL/hour]
~3 x fresh gas flow [L/min] x concentration [vol.%]
Rough formula for running time [hours] =
(for 150 mL anaesthetic agent)
Example: fresh gas flow = 2 L/min, concentration = 1.5 vol.%
Running time = 16.5 hours
Pressure range
–100 mbar to 200 mbar
Loss of anaesthetic agent when
switched off
<0.25 mL liquid/24 hours
Maximum angle of tilt, filled
45
°
(with fixed connection on anaesthetic machine)
Weight, filled
6.5 kg to 8.5 kg depending on design
Dimensions Vapor 19.n
with key-indexed filling
system (measured in mm)
Measurements in brackets apply to
Vapor 19.n with filling spout
Vapor conforms to relevant Standards
DIN 13252*
ISO 5358
BS 4272
SN 057600
CSA Z168.3*
ASTM F1161
prEN 740 (draft)*
Key-indexed filling system
ISO 5360
prEN 1280
CSA Z168.4
________________
*
Standards marked demand a key-indexed filling system
Vapor 19.3: a = 12 mm
Vapor 19.2: a = 8 mm
Vapor 19.1: a = nicht vorhanden
143 (135)
205 a
ø 106
116 (106)
Vol. %
0
Technical Data
28
50
fresh gas flow [L/min]
●
concentration [vol.%]
not available