Maintenance Intervals
Clean and disinfect Vapor before each maintenance* –
and also when returning for repair.
Inspection and service
Every six months, at the same time as the anaesthetic
machine, to be carried out by trained service personnel.
A record must be kept.
Recommendation: Call DrägerService for inspection and
service.
Packing for dispatch
Drain Vapor completely, clean and disinfect.
Each Vapor must be packed individually with care.
Use original packing, when possible.
If original packing is not available use strong packing,
with at least 5 cm bubble-wrap or impact-resistant
material.
Fasten package securely.
Taking back of Vapors
When repair is not economical, DrägerService offer to
take back Vapors for proper disposal.
___________
* According to DIN 31 051 the following definitions apply:
Inspection
= determining actual condition
Service
= measures to maintain required condition
Repair
= measures to re-establish required condition
Maintenance = inspection, service and, when necessary, repair
Maintenance Intervals
26