Setting concentration of anaesthetic agent
Only use Vapor which is in proper working order.
●
First set flow of fresh gas on anaesthetic machine,
then
1 press »0« button and
2 turn handwheel anti-clockwise to the concentration of
anaesthetic agent desired.
●
Concentrations of less than 0.2 (0.3) vol.% must
not be set.
Below these values the concentration delivered is not
defined.
●
Check filling level in sight-glass regularly and fill the
Vapor when minimum mark is reached, at the latest:
Filling Vapor, see page 9.
When changing to a Vapor for a different anaesthetic
agent:
●
Switch anaesthetic agent monitor over to the new
anaesthetic agent.
Use with scanner (MRT, NMR, NMI)
The Vapor can be used during scanning with magnetic
fields of up to 70 millitesla.
However when changing Vapor, do so outside of
scanner room.
When removing Vapor, do so outside of scanner
room.
The Vapor contains ferro-magnetic parts and will be
attracted by the magnetic force of the scanner.
●
Vapor may only be used with an anaesthesia machine
suitable for NMR tomography, such as Titus NMR.
●
Only use monitors which are suitable for use with
scanners.
Use when at an angle
e.g. with emergency equipment
Fixed Vapors may be operated at an angle of up to 45
°
.
Plug-in systems may leak when the device is in a sloping
position.
The filling level in the sight-glass will be affected by the
sloping position of the device. Place the device in vertical
position when checking filling level and filling.
If Vapor is accidentally tilted by more than 45
°
, check
concentration being delivered, see Behaviour when tilted,
page 39.
17
Operation
Setting concentration of anaesthetic agent
Use with scanner
Use when at an angle
1
2