Operation
Checks before starting anaesthesia
●
Anaesthetic agent monitor is switched on and set to
correct anaesthetic agent.*
During low-flow and minimal-flow operation, the
concentration of anaesthetic agent in the fresh gas
may be very different from the concentration in the
breathing system. It is, therefore, essential to
measure the inspiratory and/or expiratory anaesthetic
agent concentration.
●
O
2
monitor is switched on.
●
There is enough anaesthetic agent in Vapor.
●
Sealing block on key-indexed safety filling system is in
place and locking screw is tightened.
●
Filling spout: inlet and outlet valve are closed.
●
Handwheel is engaged at »0«.
●
Vapor is connected to anaesthetic machine properly
and sealed.
●
Seals on plug-in system are perfect.
●
Vapor plug-in adaptor is in a level position.
●
Locking lever turned to left.
●
It is only possible to operate one Vapor at a time.
●
Anaesthetic machine has been prepared in
accordance with the relevant Instructions for Use and
scavenging system is connected.
________________
*
The maximum adjustable alarm limit on the anaesthetic agent
monitor must be greater than Vapor's maximum concentration.
Operation
Checks before starting anaesthesia
16