![Dräger Vapor 19.n Instructions For Use Manual Download Page 22](http://html.mh-extra.com/html/drager/vapor-19-n/vapor-19-n_instructions-for-use-manual_2526843022.webp)
Care
Cleaning
Wipe heavy stains off Vapor with disposable cloth.
Do not immerse Vapor in liquid.
If other liquids get into the vaporising chamber, the
patient may be harmed, or corrosive products may be
formed, e. g. by ingression of water, and affect the
Vapor's performance.
Do not put filling adaptor in water or disinfectant
solution.
For Halothane Vapors, observe the following.
Halothane contains thymol to stabilise it, which is less
volatile than Halothane and therefore, accumulates in the
Vapor. Thymol and other reaction products may gradually
affect the functioning of the wick inside the vaporising
chamber and colour the Halothane yellow.
When dirt particles are present in the sight-glass or when
the Halothane has turned yellow:
●
Drain off discoloured Halothane (see page 19 and
21).
●
Re-fill to the maximum mark with fresh Halothane (see
page 9), allow to interact for about 6 hours and then
drain completely.
●
Dispose of drained Halothane in accordance with
regulations.
If yellow discolouration persists:
●
Have wick replaced by trained service personnel.