Vapor with filling spout
●
Keep Vapor vertical.
●
Clean filling spout with dry cloth.
1 Switch off Vapor = turn handwheel to »0«,
»0« button engaged.
When filling during operation:
●
The fresh gas flow may remain switched on.
●
After handwheel has been set to »0«, wait for
5 seconds for pressure to equalise.
If the filling system is opened straight away, the liquid
anaesthetic agent will bubble out, because the
vaporising chamber is pressurised, when the Vapor is
switched on and in operation.
●
Use the correct anaesthetic agent bottle.
2 Open inlet valve by about 3 turns.
3 Outlet valve must be closed.
●
Pour anaesthetic agent into the filling spout – note
filling level on sight glass – fill to maximum mark
only.
2 Close inlet valve completely.
●
Close anaesthetic agent bottle.
If filling has exceeded maximum mark:
●
Drain off excess anaesthetic agent,
see "Draining Vapor", page 21.
13
Preparation
Filling Vapor
0
1
2
Vol. %
3
Min.
Max.